Skip to main content

Table 3 Bias and 95% limits of agreement in non-normalized and normalized TOF-Watch SX and TOFscan

From: Comparison of the TOFscan and the TOF-Watch SX during pediatric neuromuscular function recovery: a prospective observational study

TOF-Watch SX measurements

Bias ± standard error (TOF-Watch SX – TOFscan; 95% CI)

SD ± standard error of the differences

95% Limits of agreement

95% CI, lower limit of agreement

95% CI, upper limit of agreement

Non-normalizeda

0.010±0.007

(−0.003, 0.0236)

0.127±0.0120

−0.239 to 0.259

−0.263 to −0.216

0.235 to 0.282

Normalizedb

−0.002±0.006

(−0.013, 0.010)

0.111±0.010

−0.219 to 0.216

−0.239 to −0.199

0.196 to 0.236

  1. a338 measurements in 35 individuals
  2. b345 measurements in 35 individuals