Table 3 Technical performance of patients taken to general anesthesia with closed loop TIVA technique and SEDLine monitoring

From: Technical and clinical evaluation of a closed loop TIVA system with SEDLineTM spectral density monitoring: Multicentric prospective cohort study

Care center

Institution 1: IPS Universitaria, Medellín

(n = 47)

Institution 2: Hospital San José, Bogotá

(n = 27)

Institution 3: S.E.S Hospital de Caldas

(n = 19)

Consolidated Global (n = 93)

MDPE*

11 (− 18–18)

10.7 (− 10.7–21.4)

10.7 (− 7.1–25)

10.7 (− 11–18)

MDAPE*

21 (14–25)

21.4 (4.2–32.1)

17.6 (14.2-28.5)

21 (14–26.8)

WOBBLE*

7 (7–11)

14.2 (10.7–17.8)

10.7 (7.1-14.2)

10.7 (7–16.9)

  1. MDPE, Technical median performance error
  2. MDAPE, median absolute performance error
  3. WOBBLE, within-subject variability in the infusion system. Median (interquartile range 25–75)