Table 3 Technical performance of patients taken to general anesthesia with closed loop TIVA technique and SEDLine monitoring

From: Technical and clinical evaluation of a closed loop TIVA system with SEDLineTM spectral density monitoring: Multicentric prospective cohort study

Care centerInstitution 1: IPS Universitaria, Medellín
(n = 47)
Institution 2: Hospital San José, Bogotá
(n = 27)
Institution 3: S.E.S Hospital de Caldas
(n = 19)
Consolidated Global (n = 93)
MDPE*11 (− 18–18)10.7 (− 10.7–21.4)10.7 (− 7.1–25)10.7 (− 11–18)
MDAPE*21 (14–25)21.4 (4.2–32.1)17.6 (14.2-28.5)21 (14–26.8)
WOBBLE*7 (7–11)14.2 (10.7–17.8)10.7 (7.1-14.2)10.7 (7–16.9)
  1. MDPE, Technical median performance error
  2. MDAPE, median absolute performance error
  3. WOBBLE, within-subject variability in the infusion system. Median (interquartile range 25–75)